Reliability and criterion validity of abductor hallucis muscle volume measurements using ultrasound and MRI by assessors with limited image analysis experience

Jacob B. Willes, Emma Robinson, Joshua K. Sponbeck, Derek A. Swanson, Dallin C. Swanson, Austin G. Goodrich, Steven P. Allen, A. Wayne Johnson

Abstract

The abductor hallucis (ABDH) contributes to foot function and mobility, which can influence quality of life, making accurate assessment of muscle volume important. While magnetic resonance imaging (MRI) is often considered to be the reference standard of medical imaging, ultrasound (US) provides a cost-effective, rapid and dynamic way of assessing muscle size and function.

Introduction

Foot muscles are vital in contributing to various functions of the human body [1] such as walking and are directly correlated with quality of life [2]. These muscles provide support and stability for the body during stance [3] and bipedal locomotion [4].

Materials and methods

Twenty-two participants were recruited for measurement of right and left ABDH muscle volumes measured via US and MRI (females = 12, males = 10) (Age 54.6 ± 6.63; Ht: 174.0 ± 9.3 cm; Wt: 84.5 ± 15.4 kg), a potential total of 44 feet. Of these, 7 feet were excluded due to protocol deviations or imaging artifacts.

Results

Side-specific repeatability was high across measurement methods. Right-foot ICC(3,1) values were 0.962 for MRI-OX, 0.946 for MRI-TC, and 0.997 for US-TC, while left-foot values were 0.993, 0.991, and 0.932, respectively (Table 1).

Discussion

The objective of this study was to evaluate the reliability and criterion validity of ABDH muscle volume measurements derived from US and MRI images by assessors with limited experience in image-analysis.

Conclusion

ABDH muscle volume measurements derived from US and MRI images by assessors with limited prior image-analysis experience demonstrated high repeatability, strong linear associations, and statistical equivalence within the ± 10% margin.

Citation: Willes JB, Robinson E, Sponbeck JK, Swanson DA, Swanson DC, Goodrich AG, et al. (2026) Reliability and criterion validity of abductor hallucis muscle volume measurements using ultrasound and MRI by assessors with limited image analysis experience. PLoS One 21(10): e0360725. https://doi.org/10.1371/journal.pone.0360725

Editor: Jeremy P. Loenneke, University of Mississippi, UNITED STATES OF AMERICA

Received: May 12, 2026; Accepted: September 30, 2026; Published: October 21, 2026

Copyright: © 2026 Willes et al. This is an open access article distributed under the terms of the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author and source are credited.
Data Availability: All relevant data are within the paper.

Funding: The author(s) received no specific funding for this work.

Competing interests: The authors have declared that no competing interests exist.